电脑沥青针入度仪技术指导
2019-08-01 10:10:34 来源:荣计达
一,电脑沥青针入度仪产品简介:
新型电脑沥青针入度仪是一种采用数字式位移传感器,结合微电脑技术研制的测量沥青针入度测量仪器,该仪器满足GB4509-84《石油沥青针入度的测量法》,ASMD5-83/IP49-79《沥青针入度的测量法》和TO60-93《沥青针入度试验》。该仪器使用简单、方便、低温漂,带有自动控温,LCD中文显示,实时时钟等功能,适用环境范围大,实现了试验过程自动化、数字化、电脑化、为我国公路、市政、建筑、石油等领域提供了一种精度高、稳定度好、性能好的电脑自动化沥青试验仪器。
the product introduction:
The new computer asphalt penetration instrument is a measuring instrument for measuring asphalt penetration using advanced digital displacement sensor combined with microcomputer technology. The instrument meets GB4509-84 "Measurement method of penetration of petroleum asphalt", ASMD5- 83/IP49-79 "Asphalt Penetration Measurement Method" and TO60-93 "Asphalt Penetration Test". The instrument is simple, convenient, low temperature drift, with automatic temperature control, LCD Chinese display, real-time clock and other functions, suitable for a wide range of environments, to achieve automation, digitization, computerization of the test process, for China's highway, municipal, construction, petroleum The field provides a computerized automated asphalt test instrument with high precision, good stability and reliable performance.
二,电脑沥青针入度仪技术参数:
★小分度:位移:0.01mm 时间:0.1s 温度:0.1℃
★大量程:40mm
★针连杆、砝码、标准针合重:100±0.05g 150±0.05g
★控温精度:0.1℃
★电源:AC220V±10%,50HZ
★试验皿:金属制,平底状,尺寸为70×45 55×35(mm)
★平底保温皿:玻璃制品,容量大于1.5L,皿深大于60mm;附放试样皿用三角架一个。
★外形尺寸:250X220X700mm
★重量:15kg
Second, technical parameters:
★Minimum indexing: Displacement: 0.01mm Time: 0.1s Temperature: 0.1°C
★Maximum range: 40mm
★ Needle connecting rod, weight, standard needle combined weight: 100±0.05g 150±0.05g
★ Temperature control accuracy: 0.1 °C
★Power supply: AC220V±10%, 50HZ
★Testing dish: metal, flat bottom, size 70×45 55×35 (mm)
★ Flat-bottom insulation dish: glassware, the capacity is more than 1.5L, the depth of the dish is more than 60mm; the sample holder is attached with a tripod.
★Dimensions: 250X220X700mm
★ Weight: 15kg
三,电脑沥青针入度仪使用方法:
★将预先脱水的试样加热溶化,加热温度不得高于试样估计软化点90℃,时间不得多于30min,充分搅拌过筛后注入盛样皿内。其深度至少应超过预期针入度值约10mm放置于15-39℃的室温中冷却1h,冷却时应注意不应有灰尘落入。
★将盛样 皿浸入盛满清水的水槽中,保温1-1.5h,水温保持25±0.5℃,水面应高于试验表面25mm以上。
★调整针入度仪,使之水平(必要时调节附加砝码的重量)用手揿下释放按钮,提起标准针针杆,使针杆顶端面与百分表接触头接触。
★盛样皿保温1-1.5h后供应商河北大宏实验仪器有限公司,取出并放入水温严格控制为25℃的平底保温皿中,试样表面以上的水层高度应不少于10mm,将保温皿置于针入度仪的圆形工作台上,调节工作台,使试样表面与针尖恰好接触(如调节工作台不够,可用调节螺母调节杠架与之配合),将百分表指针对零。
★将时间开关拨至试样所需的时间(5s或60s)档位上,按一下开始按钮,电磁铁吸合使标准针自由地穿入试样中,至红灯亮时(此时,表示标准针穿入试样中的时间已达5s或60s,电磁铁自动放磁弹簧将针杆固紧,使其针杆不再下落),用手按下百分表触头与标准针针外杆顶端面接触,此时百分表指针所指的数,即为试样的针入度。
★同一试样重复试验至少三次,每次穿入点的相互距离及与盛样皿边缘的距离都不得小于10mm,各次试验结果的平均值即为此试样的试验结果。
★每次试验前,都应检查并调节保温皿内的水温,试验后都应将标准针取下用浸有溶剂(汽油)的棉花或布拭净,再用棉花或布擦干。
★ The pre-dehydrated sample is heated and melted. The heating temperature should not be higher than the estimated softening point of the sample by 90 ° C, the time should not exceed 30 min, and the mixture is fully stirred and sieved into the sample dish. The depth should be at least 10 mm beyond the expected penetration value and placed at room temperature of 15-39 ° C for 1 h. When cooling, care should be taken that no dust should fall.
★ Immerse the sample in a water tank filled with clean water for 1–1.5 hours, keep the water temperature at 25±0.5°C, and the water surface should be 25mm or more above the test surface.
★Adjust the needle penetration meter to make it level (adjust the weight of the additional weight if necessary). Press the release button by hand and lift the standard needle bar so that the top end of the needle bar contacts the contact lens of the dial indicator.
★ After the sample is insulated for 1-1.5h, the supplier Hebei Dahong Experimental Instrument Co., Ltd. shall be taken out and placed in a flat-bottomed insulated dish with a water temperature strictly controlled at 25 °C. The height of the water layer above the surface of the sample shall be not less than 10 mm. Place the insulated dish on the circular table of the penetrometer and adjust the table so that the surface of the sample is in good contact with the tip (if the table is not enough, the adjusting nut can be used to adjust the bar frame). Pointer to zero.
★ Put the time switch to the time required for the sample (5s or 60s), press the start button, and the electromagnet pulls the standard needle freely into the sample until the red light is on (at this time, Indicates that the standard needle has penetrated into the sample for 5s or 60s. The electromagnet automatic demagnetization spring tightens the needle bar so that the needle bar no longer falls. Press the dial indicator and the standard needle by hand. The top end of the outer rod is in contact, and the number indicated by the dial indicator is the penetration of the sample.
★ Repeat the test at least three times for the same sample. The mutual distance between each penetration point and the distance from the edge of the sample dish should not be less than 10mm. The average value of each test result is the test result of this sample.
★ Before each test, the water temperature in the incubator should be checked and adjusted. After the test, the standard needle should be removed and wiped with cotton or cloth soaked with solvent (gasoline), and then dried with cotton or cloth.
四,电脑沥青针入度仪安装与调试
★新仪器开箱后,应按装箱单检查配件是否齐全,有无损坏。将针入度仪主机放在水平的平台上,接通电源,按住<手动>键使电磁铁吸合,将标准针的连杆插入传感器的孔内,放开<手动>键。把标准针与连杆联接即可。
★开箱后,检查是否因运输而发生故障。
★按装箱清单检查是否缺件,及时通知本公司销售部或销售商。
★将电源插上,打开电源,此时显示屏显示仪器名称后进入试验界面。
★为确保检测精度,通电15S后,按下仪器的<手动>键,此时针杆组件应能自由落下。
★调整仪器水平方法如下:转动调节支角,使水平泡中气泡处于中心位置,此时,工作平面已处于水平位置。
Fourth, installation and commissioning
★After unpacking the new instrument, check whether the accessories are complete and damaged according to the packing list. Place the needle gauge on the horizontal platform, turn on the power, press and hold the button to pull the electromagnet, insert the pin of the standard pin into the hole of the sensor, and release the button. Connect the standard needle to the connecting rod.
★ After unpacking, check if there is a malfunction due to transportation.
★ Check the missing parts according to the packing list and notify the sales department or seller of the company in time.
★ Plug in the power and turn on the power. At this time, the display shows the instrument name and enters the test interface.
★ To ensure the detection accuracy, after the power is turned on for 15 seconds, press the button of the instrument, and the needle bar assembly should be free to fall.
★ The method of adjusting the level of the instrument is as follows: Turn the adjusting angle to make the bubble in the horizontal bubble in the center position. At this time, the working plane is in the horizontal position.